OPTITEMP TCA-TS36 – Standard version
search
  • OPTITEMP TCA-TS36 – Standard version

OPTITEMP TCA-TS36

OPTITEMP TCA-TS36

Thermocouple (TC) temperature assembly for higher flow velocities and pressures

Thread connection

DIN-style based on form 7, barstock thermowell with straight tip

-40…+600°C / +1100°F

 

Shipping Policy

 

Return policy

 

Terms & Conditions

The OPTITEMP TCA-TS36 is a threaded thermocouple (TC) assembly with a straight barstock thermowell (based on form 7). The temperature assembly for advanced requirements is used when the strength of welded multipart thermowells is not sufficient and a high mechanical stress resistance of the thermowell is required. It is therefore particularly suitable for temperature measurement in pipes, tanks or reactors at higher flow velocities and pressures (e.g. in the petrochemical or oil and gas industry). The TC sensor is designed for process temperatures up to +600°C / +1100°F.

The OPTITEMP TCA-TS36 features a modular design with a threaded standard neck tube that can be exchanged easily, even under operating conditions. This makes the temperature assembly very flexible. If it has been damaged from the outside, servicing can be carried out without process shut down. The thermocouple assembly can be combined with most thermowells. It is available with ATEX and IECEx approvals.

Features & Options

  • Temperature measurement of liquids, gases and steam
  • Modular design: threaded standard neck tube for easy exchange under operating conditions
  • Thermowell material: 1.4404 / 316L or 1.4571 / 316Ti
  • Standard or customised lengths
  • Different connection heads available (IP54…IP68)
  • With optional temperature transmitter
  • Available as intrinsically safe (Ex i) and flameproof (Ex d): ATEX, IECEx

Typical applications

  • Applications with advanced requirements
  • Temperature measurement of media requiring high mechanical stress resistance
  • Suitable for applications where assemblies complying with DIN standards are required
  • Primarily used in the petrochemical industry, the oil and gas industry as well as in industries such as:
    • Chemical
    • Power generation
    • Machinery and apparatus / OEM
    • Iron and steel
    • Pulp and paper

Specifications

Depending on selected device configuration

General information

Product type Temperature assembly
Product family name OPTITEMP
Product positioning For higher flow velocities and pressures
Product variants
  • Standard: OPTITEMP TCA-TS36
  • EX: OPTITEMP TCA-TS36 Ex
Measurement variables Temperature
Measurement principle Thermocouple (TC)
Measured media Gases, Liquids, Steam
Assembly with thermowell Yes
Temperature assembly design DIN-style
Thermowell type Barstock
Thermowell form (according to DIN 43772) Form 6, Form 7
Sensor mounting Screw-in
Measuring insert replaceable Yes
Measuring insert type Replaceable spring loaded mineral isolated measuring insert OPTITEMP TC 100
Tolerance class (according to DIN EN 60584) 1
Temperature sensor connection type option Ceramic terminal block, Flying wires, Temperature transmitter
Thermocouple (TC) configuration 1 x Type J, 1 x Type K, 2 x Type J, 2 x Type K
Thermowell diameters Ø17 mm / 0.67"
Thermowell tip design Straight
Mounting point Pipe
Measuring point Pipe

Operating conditions

Process temperature -40…600°C/
-40…1100°F

Materials

Wetted parts Austenitic stainless steel (1.4404/316L), Austenitic stainless steel (1.4571/316Ti)

Process connections

Thread connections Male:
G½, G¾, G1
½ NPT, ¾ NPT

Approvals/Standards

Certificates/Approvals Ex ATEX, IECEx

OPTITEMP TCA-TS36
Comments (0)